Analytical-industries AII-2000 M Oxygen Analyzer Manual de usuario

Busca en linea o descarga Manual de usuario para Equipo Analytical-industries AII-2000 M Oxygen Analyzer. Analytical Industries AII-2000 M Oxygen Analyzer User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 18
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Copyright © 10/10 All Rights Reserved
Analytical Industries Inc.,
2855 Metropolitan, Pomona, CA 91767 USA.
Tel: 909-392-6900, Fax: 909-392-3665
e-mail: sales-medical@aii1.com, web: www.aii1.com
This manual may not be reproduced in whole or in part without
the prior written consent of Analytical Industries Inc.
510(k) K952736 O2 Sensor
510(k) K053407 O2 Analyzers
EN 12598:1999 / ISO 7767: 1997
(Oxygen monitors for patient breathing
mixtures – Particular requirements).
EC: Medical Device Directive: 93/42/EEC,
Annex II as Amended by 2007/47/EC,
Certificate #485CE
ISO 9001:2008, Certificate #485
Authorized EC Representative:
RGV Lda.
Rua Jose Joaquim de Freitas, 247
2750-404 Cascais-Portugal
Health Canada: Medical Device Regula-
tions, F-27/ SOR-98-282
ISO 13485:2003, Certificate #485B
AII-2000 HC
AII-2000 A
AII-2000 M
Vista de pagina 0
1 2 3 4 5 6 ... 17 18

Indice de contenidos

Pagina 1

Copyright © 10/10 All Rights Reserved Analytical Industries Inc., 2855 Metropolitan, Pomona, CA 91767 USA. Tel: 909-392-6900, Fax: 909-392-3665 e-ma

Pagina 2 - 1 Introduction

Procedure AII 2000 Series Analyzers and Monitor employ the identical calibration routine and displays but they differ slightly in the way they arrive

Pagina 3

4.8 Mounting Every analyzer and monitor is equipped with a male dove tail bracket and triangular shaped thick metal wire stand secured to the rear of

Pagina 4

5.2 Application Considerations Effect of Anesthetic Agents The AII 2000 Series Oxygen Analyzers and Monitors utilize an electrochemical galvanic fue

Pagina 5 - 3 Safety Warnings

Calibrate at the temperature and pressure (altitude) at which the ana-lyzer will be operated, humidified gases cannot be 100% oxygen. Effect of Cond

Pagina 6 - 4 Start-Up

5.4.3 AII 2000HC Oxygen Analyzer (Integral Oxygen Sensor) AII 2000HC with its integral oxygen sensor requires connecting the ¼” tubing supplied (sect

Pagina 7

6.3.2 Procedure AII 2000HC - Integral Sensor 1. Tools required: small bladed screwdriver. 2. Place the device face down on a flat surface. 3. Remo

Pagina 8

28 29 Symptom Corrective Action Reading displayed by LCD drifts during calibration Wait 5 minutes and repeat calibration with sensor placed on

Pagina 9

8 Specifications 30 Expected Sensor Life Considers the full range of the sensor’s signal, example 7-13 mV. Most medical oxygen sensors are configured

Pagina 10

32 9 Warranty Coverage Under normal operating conditions, the analyzer and sensors are warranted to be free of defects in materials and workmanship

Pagina 11 - 5 Operation

Table of Contents 1 Introduction 1.1 Indications for Use 1.2 Intended Use 1.3 Device Description 1.4 Declaration of

Pagina 12

The device is intended to be re-usable. Should the device or accessories come in contact with patient bodily fluids, either dispose of the device or c

Pagina 13

1.4 Declaration of Conformity We hereby declare the above product meets the provisions of the directives and standards specified. All supporting doc

Pagina 14 - 6 Maintenance

3 Safety Warnings ALWAYS follow the statements below as they are essential to reduc-ing the risk of use error due to ergonomic features of the device

Pagina 15 - 7 Troubleshooting

4 Start-Up 4.1 Contents of Shipping Container: 4.1.1 AII 2000 A Oxygen Analyzer, and AII 2000 M Oxygen Monitor: ENCL-1061 V-mou

Pagina 16

4.3 Install Oxygen Sensor The device cannot function until the oxygen sensor is installed. Once installed, allow the sensor to stabilize for 15-20 mi

Pagina 17 - 8 Specifications

4.5 Start-Up Test Press the ON/OFF key on the front panel to apply power to the device and initiate a complete diagnostic test of all system function

Pagina 18 - 9 Warranty

4.6 Alarms AII 2000M Oxygen Monitor The monitor is equipped with user selectable HI and LO alarm set points which are displayed at the bottom of the

Comentarios a estos manuales

Sin comentarios