Procedure AII 2000 Series Analyzers and Monitor employ the identical calibration routine and displays but they differ slightly in the way they arrive
4.8 Mounting Every analyzer and monitor is equipped with a male dove tail bracket and triangular shaped thick metal wire stand secured to the rear of
5.2 Application Considerations Effect of Anesthetic Agents The AII 2000 Series Oxygen Analyzers and Monitors utilize an electrochemical galvanic fue
Calibrate at the temperature and pressure (altitude) at which the ana-lyzer will be operated, humidified gases cannot be 100% oxygen. Effect of Cond
5.4.3 AII 2000HC Oxygen Analyzer (Integral Oxygen Sensor) AII 2000HC with its integral oxygen sensor requires connecting the ¼” tubing supplied (sect
6.3.2 Procedure AII 2000HC - Integral Sensor 1. Tools required: small bladed screwdriver. 2. Place the device face down on a flat surface. 3. Remo
28 29 Symptom Corrective Action Reading displayed by LCD drifts during calibration Wait 5 minutes and repeat calibration with sensor placed on
8 Specifications 30 Expected Sensor Life Considers the full range of the sensor’s signal, example 7-13 mV. Most medical oxygen sensors are configured
32 9 Warranty Coverage Under normal operating conditions, the analyzer and sensors are warranted to be free of defects in materials and workmanship
Table of Contents 1 Introduction 1.1 Indications for Use 1.2 Intended Use 1.3 Device Description 1.4 Declaration of
The device is intended to be re-usable. Should the device or accessories come in contact with patient bodily fluids, either dispose of the device or c
1.4 Declaration of Conformity We hereby declare the above product meets the provisions of the directives and standards specified. All supporting doc
3 Safety Warnings ALWAYS follow the statements below as they are essential to reduc-ing the risk of use error due to ergonomic features of the device
4 Start-Up 4.1 Contents of Shipping Container: 4.1.1 AII 2000 A Oxygen Analyzer, and AII 2000 M Oxygen Monitor: ENCL-1061 V-mou
4.3 Install Oxygen Sensor The device cannot function until the oxygen sensor is installed. Once installed, allow the sensor to stabilize for 15-20 mi
4.5 Start-Up Test Press the ON/OFF key on the front panel to apply power to the device and initiate a complete diagnostic test of all system function
4.6 Alarms AII 2000M Oxygen Monitor The monitor is equipped with user selectable HI and LO alarm set points which are displayed at the bottom of the
Comentarios a estos manuales